Range: | -999~999mV | Deepest Depth: | Underwater 30m |
---|---|---|---|
Protection Level: | IP68 | Maximum Operating Pressure: | 6bar |
Output Signal: | RS-485, MODBUS Protocol | Power Information: | DC 5~12V, Current <50mA |
Probe Cable Length: | 5 Meters (default), Can Be Customized | Housing Material: | POM |
Assembly: | Input Type | Temperature Range: | 0~ 50°C |
High Light: | RS485 Oxidation Reduction Potential Sensor,Digital Oxidation Reduction Potential Sensor,Online ph and orp sensor |
item
|
value
|
Warranty
|
1 year
|
Customized support
|
OEM, ODM, OBM
|
Place of Origin
|
China
|
Brand Name
|
Xi'an Desun
|
Range
|
-999~999mV
|
Deepest depth
|
Underwater 30m
|
Protection level
|
IP68
|
Maximum operating pressure
|
6bar
|
Output signal
|
RS-485, MODBUS protocol
|
Power information
|
DC 5~12V, current <50mA
|
Probe cable length
|
10 meters (default), can be customized
|
Housing material
|
POM
|
Assembly
|
Input type
|
Temperature range
|
0~ 50°C
|
Application examples of orp sensor
The probe contains sensitive optical and electronic components. Ensure that the probe is not subject to severe mechanical impact. Damage to the sensor due to impact or manpower will not be covered by the warranty. Water ingress to the sensor due to replacement of the ORP electrode is not covered by the warranty.
Error | Possible causes | Solutions |
The operation interface cannot be connected or the measurement results are not displayed | Controller and cable connection error | Reconnect the controller and cables |
Cable failure | Please contact us | |
Measured values are too high, too low, or the values remain unstable | Measurement shifted | Recalibration |
Water has entered the product | Please contact us | |
The ORP electrolyte is exhausted | Replace the ORP electrochemical part |